Prospective randomized trial comparing alumina ceramic-on-ceramic with ceramic-on-conventional polyethylene bearings in total hip arthroplasty

J Arthroplasty. 2010 Apr;25(3):392-7. doi: 10.1016/j.arth.2009.01.013. Epub 2009 Feb 5.

Abstract

This prospective randomized study aims to compare the outcome between an alumina ceramic-on-ceramic (CC) articulation with a ceramic on ultra-high-molecular-weight polyethylene articulation (CP). Fifty-six hips in 55 patients with mean age 42.2 (range, 19-56) each received uncemented components, a 28-mm alumina head with randomization of acetabular liner. Mean St Michael's outcome score for each group with up to 10 years follow-up (median, 8 years; range, 1-10) was 22.8 and 22.9, respectively (P = .819). Wear was identified in all but 1 CP replacement, but only 12 of the 23 CC. Mean wear in the CP group was 0.11 mm/y and 0.02 mm/yr in the CC group (P < .001). Other than significantly greater wear in the polyethylene group, there was no significant difference in midterm outcome between the 2 groups.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Adult
  • Aluminum Oxide*
  • Arthroplasty, Replacement, Hip / instrumentation
  • Arthroplasty, Replacement, Hip / methods*
  • Bone Diseases, Developmental / surgery*
  • Ceramics*
  • Female
  • Follow-Up Studies
  • Hip Joint / diagnostic imaging
  • Hip Joint / surgery
  • Hip Prosthesis*
  • Humans
  • Male
  • Middle Aged
  • Osteoarthritis, Hip / surgery*
  • Osteonecrosis / surgery
  • Polyethylene*
  • Prospective Studies
  • Radiography
  • Treatment Outcome

Substances

  • Polyethylene
  • Aluminum Oxide