Diagnosis of type III endoleak and endovascular treatment with aortouniiliac stent-graft

J Vasc Interv Radiol. 2009 Jan;20(1):125-9. doi: 10.1016/j.jvir.2008.09.029. Epub 2008 Nov 20.

Abstract

The present report describes a case of type III endoleak from a tear in the fabric of a Zenith bifurcated stent-graft approximately 6 months after implantation. The reason of the fabric tear was unknown. The complication was successfully treated by aortouniiliac stent-graft implantation followed by creation of a femorofemoral bypass.

Publication types

  • Case Reports

MeSH terms

  • Angiography, Digital Subtraction
  • Aortic Aneurysm, Abdominal / diagnostic imaging
  • Aortic Aneurysm, Abdominal / surgery*
  • Aortography / methods
  • Balloon Occlusion
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Blood Vessel Prosthesis Implantation* / instrumentation
  • Blood Vessel Prosthesis*
  • Humans
  • Iliac Artery / diagnostic imaging
  • Iliac Artery / surgery*
  • Male
  • Middle Aged
  • Postoperative Complications / diagnostic imaging
  • Postoperative Complications / surgery*
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation
  • Stents*
  • Tomography, X-Ray Computed