Late type III endoleak from fabric tears of a zenith stent graft: report of a case

  • Case report
  • Published:
Surgery Today Aims and scope Submit manuscript

Abstract

We report a case of late type III B endoleak from a fabric tear in the main body of a Zenith bifurcated stent-graft approximately 19 months after implantation. A follow-up computed tomography (CT) scan showed gradual growth in the size of the aneurysm, but no apparent endoleak was recognized. However, the CT scan had demonstrated a mosaic pattern in the aneurysm sac, which could indicate an endoleak from a fabric tear. The defect was repaired by implantation of body extensions, but the patient died 5 h after the procedure.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Parodi JC, Palmaz JC, Barone HD. Transfemoral intraluminal graft implantation for abdominal aortic aneurysms. Ann Vasc Surg. 1991;5(6):491–9.

    Article  PubMed  CAS  Google Scholar 

  2. White GH, Yu W, May J. Endoleak—a proposed new terminology to describe incomplete aneurysm exclusion by an endoluminal graft. J Endovasc Surg. 1996;3(1):124–5.

    Article  PubMed  CAS  Google Scholar 

  3. White GH, Yu W, May J, Chaufour X, Stephen MS. Endoleak as a complication of endoluminal grafting of abdominal aortic aneurysms: classification, incidence, diagnosis, and management. J Endovasc Surg. 1997;4(2):152–68.

    Article  PubMed  CAS  Google Scholar 

  4. White GH, May J, Waugh RC, Chaufour X, Yu W. Type III and type IV endoleak: toward a complete definition of blood flow in the sac after endoluminal AAA repair. J Endovasc Surg. 1998;5(4):305–9.

    Article  PubMed  CAS  Google Scholar 

  5. Teutelink A, van der Laan MJ, Milner R, Blankensteijn JD. Fabric tears as a new cause of type III endoleak with Ancure endograft. J Vasc Surg. 2003;38(4):843–6.

    Article  PubMed  Google Scholar 

  6. Brown KE, Heyer KS, Matsumura JS, Eskandari MK. Late Type III endoleak and graft failure of an Ancure stent-graft. J Vasc Interv Radiol. 2008;19(10):1506–8.

    Article  PubMed  Google Scholar 

  7. Lee WA, Huber TS, Seeger JM. Late type III endoleak from graft erosion of an Excluder stent graft: a case report. J Vasc Surg. 2006;44(1):183–5.

    Article  PubMed  Google Scholar 

  8. Wanhainen A, Nyman R, Eriksson MO, Björck M. First report of a late type III endoleak from fabric tears of a Zenith stent graft. J Vasc Surg. 2008;48(3):723–6.

    Article  PubMed  Google Scholar 

  9. Juszkat R, Staniszewski R, Zarzecka A, Majewski W. Diagnosis of type III endoleak and endovascular treatment with aortouniiliac stent-graft. J Vasc Interv Radiol. 2009;20(1):125–9.

    Article  PubMed  Google Scholar 

  10. Hori D, Yuri K, Nemoto K, Yamaguchi A, Adachi H. Successful treatment of an abdominal aortic aneurysm by endovascular graft placement through a previously placed prosthetic graft: report of a case. Surg Today. 2010;40(12):1169–72.

    Article  PubMed  Google Scholar 

  11. Cao P, De Rango P, Verzini F, Parlani G. Endoleak after endovascular aortic repair: classification, diagnosis and management following endovascular thoracic and abdominal aortic repair. J Cardiovasc Surg (Torino). 2010;51(1):53–69.

    CAS  Google Scholar 

  12. Stavropoulos SW, Charagundla SR. Imaging techniques for detection and management of endoleaks after endovascular aortic aneurysm repair. Radiology. 2007;243(3):641–55.

    PubMed  Google Scholar 

Download references

Conflict of interest

Hiroshi Banno and his co-authors have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hiroshi Banno.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Banno, H., Morimae, H., Ihara, T. et al. Late type III endoleak from fabric tears of a zenith stent graft: report of a case. Surg Today 42, 1206–1209 (2012). https://doi.org/10.1007/s00595-012-0320-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00595-012-0320-8

Keywords

Navigation